Biosystems PhD Ultra Manual de instrucciones Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 60
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
23
2.4.2.2 Protective “low flow” shut-downs
If a protective pump shut-down occurs, the following
steps should be taken before the instrument is put
back into use:
(1) Turn off the PhD Ultra detector and disconnect
the sample draw pump.
(2) Remove the sample draw assembly from the
area being monitored. Be careful to keep the
sample draw probe in a vertical position.
(3) Examine the sample draw probe and hose to
make sure no fluids remain trapped.
(4) Allow any trapped fluids to completely drain. (It
may be necessary to disconnect the hose or
sample draw probe before drainage can occur.)
(5) Replace the sample draw probe filters if
necessary.
(6) Re-attach the pump in fresh air and wait for
readings to stabilize.
(7) Resume sampling.
2.4.2.3 Resuming diffusion monitoring
In order to stop using the pump and resume
diffusion monitoring, simply disconnect the pump
assembly from the PhD Ultra. The audible and
visual alarms will be activated and the LCD will
display the message “Pump Disconnect”. Press the
mode button to acknowledge and resume normal
diffusion operation.
2.4.3 Sample probe assembly
The sample probe handle contains moisture barrier
and particulate filters designed to remove
contaminants that may cause harm to either the
continuous pump or the instrument. Never operate
the sample draw pump unless the hose and
probe assembly is attached!
Particulate contaminants are removed by means of
a cellulose filter similar to those used in filter
cigarettes. The hydrophobic filter includes a 0.1 µm
Teflon barrier which blocks the flow of moisture as
well as any remaining particulate contaminants.
Figure 2.4.3. PhD Ultra sample probe assembly
Sample probe filters should be replaced whenever
visibly discolored due to contamination, or when the
continuous pump “low flow” alarm indicates
blockage. A spare filter replacement kit (part
number 54-05-K0401) is included with every PhD
Ultra.
2.4.3.1 Changing sample probe filters
The threaded sample probe handle is unscrewed (as
shown in Figure 2.4.3.) to access the filters. The
particulate filter is held in place by means of a clear
filter bowl. To replace the particulate filter, remove
the old filter and bowl, insert a new filter into the
bowl, and slide the bowl back into place in the probe
handle. The hydrophobic barrier filter fits into a
socket in the rear section of the probe handle. (The
narrow end of the hydrophobic barrier filter is
inserted towards the rear of the handle.)
2.4.3.2 Changing sample probe tubes
The standard 11.5” long butyrate probe tube is held
in place by means of a hex-nut compression fitting
and compression sleeve. The standard probe tube
is designed to be easily interchangeable with other
custom length sections of 1/4” OD tubing, or probe
tubes made of other materials (such as stainless
steel).
Probe tubes are exchanged by loosening the hex-
nut compression fitting, removing the old tube,
sliding the compression sleeve into place around the
Vista de pagina 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 59 60

Comentarios a estos manuales

Sin comentarios