92. Slide the open end of the tubing assembly over the regulator. 3. Connect the end of the tubing with the white quick disconnect fitting to the
102.7.4 Multiple Cylinders to Multiple Docks The PHD6 includes 4 calibration gas inlet ports to provide calibration flexibility for gas detection pro
11 The software will notify you when the process is complete. Click “Finish”. 2.8.2 Dock screens Once the connection is made, the dock will display
12Indicators section of the IQ Ethernet Instructions Booklet. The current status of each dock is shown in the left column. When there are no detecto
13of the screen. Once the dock configuration is uploaded, the “Set Config” button will be enabled. Once changes are made, click on the “Set Config”
14 To darken or lighten the display on the dock, move the slider as appropriate. The dock will not show the new LCD setting until the “Set Config” b
15 Once the gas type is selected, use the arrows to the right of the gas value input box to increase or decrease the value. If changes need to be ma
16 To view the dock’s Display, click on “Get Current LCD Display”. Ethernet Controls For docks that are connected by Ethernet, there are three addit
17appear on the right. Click on “View Session Data” to directly access session information for the instrument using the BioTrak Program. If the doc
18must then be programmed with the same PIN and set to check for the code in the instruments. 3. Select the dock that will be used to write the PIN
1 THE IQ6 DOCKING STATION IS DESIGNED TO INTERFACE WITH PHD6 GAS DETECTORS. SPERIAN GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION OF DEFICIEN
19 Enter your User ID and password again when the Security Login window is shown. A window will be shown to indicate that the dock configuration has
204.1 Test Sequence Once the instrument has been recognized, the fresh air / zero calibration will begin automatically. Performing the fresh air
21connected to the PC, the fault will also be indicated in the output box on the instrument tab in the IQ Express software. 4.3.1 No gas detected If
225.2.1 Dock The Dock Tab under Tools / Options includes settings for the Dock Timeout and the Calibration Gas Lot Number. Each dock has its own set
23large number of instruments are in use that are bump-tested daily though the IQ Express Dock. 5.2.4 Session Downloading The session downloading o
24CAUTION Interfering with the flash upload may result in damage to the IQ6 Dock . Once the upload is completed, you’ll need to restart the PC for t
25Appendix A: Calibration Frequency One of the most common questions that we are asked at Sperian Instrumentation is: “How often should I calibrate
26Sperian Instrumentation Warranty Gas Detection Products General Sperian Protection Instrumentation, LLC (hereafter Sperian) warrants gas detectors,
2Table of Contents Warnings and Cautions 3 A. Signal Words 3 1.Overview 4 1.1 Tests and record keeping 4 1.2 PC Connection 4 1.2.1 USB Connec
3Optimize O2 Bump Test 15 Saving new settings 15 Dock LCD Display 15 Ethernet Controls 16 3.2.2 PHD6 Tab 16 3.2.2.1 Instrument Tab 17 General
41. Overview The IQ6 Docking Station is an automatic calibration station for use with the Sperian PHD6 Gas Detectors. The IQ6 Docking Station automa
5 30MB hard drive disk space. USB Port *Sperian Instrumentation discourages multiple IQ Express Dock setups with PC’s running Windows 98. **Servi
6the docks by connecting them to the PC’s USB port. If the instrument is to be connected via Ethernet, follow the Ethernet Instructions that came wi