Biosystems PhD Ultra Manual de usuario

Busca en linea o descarga Manual de usuario para Detectores de humo Biosystems PhD Ultra. Biosystems ToxiLtd® Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 20
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
0
Biosystems
ToxiLtd
®
Single Gas
Detector
Reference Manual
Sperian Instrumentation
651 South Main Street
Middletown, CT 06457
(800) 711-6776
(860) 344-1079
Fax (860) 344-1068
Part number 13-266-N
Version 3.00
1Nov2009
Vista de pagina 0
1 2 3 4 5 6 ... 19 20

Indice de contenidos

Pagina 1 - Biosystems

0 Biosystems ToxiLtd® Single Gas Detector Reference Manual Sperian Instrumentation 651 South Main Street Middletown, CT 0

Pagina 2

9The MAX screen will be shown for 5 seconds. The instrument will then automatically return to the months or days remaining screen (unless an alarm c

Pagina 3 - Table of Contents

10it will automatically launch the IrDA connection subroutine so that data in the instrument may be downloaded to BioTrak. 2.5 Turning the ToxiLtd o

Pagina 4 - Warnings and Cautions

11calibration as discussed below in section 3.3. The fresh air/zero calibration takes less than ten seconds to perform and does not require an exter

Pagina 5

12button. Note: The maximum reading value (MAX) will be automatically reset following a successful fresh air/zero calibration. Fresh air/zero cali

Pagina 6 - 1. Overview

13  2. Apply calibration gas as shown above in figure 3.4. Once calibration gas is detected, the readout will change to show the gas reading.

Pagina 7

14warning symbol and the calibration bottle icon will be shown, which signifies that the instrument failed the last attempt to span calibrate. 3.6.2

Pagina 8 - 2. Field Operation

154.1 Initiating communications To initiate communications, simply hold the MODE button as if you are turning the instrument off. The ToxiLtd will b

Pagina 9

16Appendices Appendix A: Sensor Cross-Sensitivity Chart The table below provides the cross-sensitivity response of the ToxiLtd toxic gas sensors to c

Pagina 10

17Appendix C: Calibration Frequency One of the most common questions that we are asked at Sperian Instrumentation is: “How often should I calibrate m

Pagina 11 - 3. Calibration

18Sperian Instrumentation Warranty Gas Detection Products General Sperian Protection Instrumentation, LLC (hereafter Sperian) warrants gas detectors,

Pagina 12 - Accuracy of

1 THE BIOSYSTEMS TOXILTD-NGSG® PERSONAL PORTABLE GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION AND MEASUREMENT OF POTENTIALLY HAZARDOUS ATMOS

Pagina 13

19 ToxiLtd® c-UL Version ATEX / CQST Version IECEx Version Gas CO H2S Firmware Ver. _______ Vibrating Motor Belt Clip Serial # _

Pagina 14

2Table of Contents Certifications ... 3 Warnings and Cautions ...

Pagina 15 - 4. Event logging

34.1 Initiating communications...15 Appendices...

Pagina 16

4B. Warnings 1. The ToxiLtd personal, portable gas detectors have been designed for the detection of either oxygen deficiencies or specific toxic g

Pagina 17 - Appendices

51. Overview The Biosystems ToxiLtd hereafter referred to as the ToxiLtd is a single sensor gas detector that is designed to detect either carbon mon

Pagina 18 - Frequency

6 In the event of an alarm condition it is important to follow established procedures. The safest course of action is to immediately leave the affe

Pagina 19 - General

72. Field Operation Field operation of the ToxiLtd is controlled entirely through the MODE button, which is located on the front of the instrument.

Pagina 20 - ToxiLtd

8The sensor type will be shown next. During the display of the warning alarm level, the LED alarm light will flash twice and the audible warning ala

Comentarios a estos manuales