The sensor type will be shown next.
During the display of the warning alarm
level, the LED alarm light will flash twice
and the audible warning alarm will sound
twice.
During the display of the danger alarm
level, the LED visual alarm will flash twice
and the audible danger alarm will sound
twice.
Following the danger alarm level, the
ToxiLtd will begin standard operation
unless calibration is due otherwise it will
display either the Gas Indicator screen if
enabled or it will show the number of
months remaining in the 24-month
lifecycle of the instrument. During an
alarm condition, the instrument will display
the current gas reading.
For the first full month of use following
initialization the display will show 23 full
months remaining.
Once the instrument determines that there
are 90 days of operation remaining, the
instrument will indicate 90 days and
proceed to count the days down until the
instrument reaches 0 days and turns itself
off.
2.2 Functions
As stated above, the ToxiLtd will show the
Gas Indicator screen if enabled or it will
show the amount of time remaining in the
product’s lifecycle during normal
operation unless there is an alarm
present. The ToxiLtd will display gas
readings while it is in alarm.
2.2.1 View time remaining, max gas
readings, time in alarm, calibration due
If the Gas Indicator screen is showing
press the MODE button once to view the
time remaining screen. From the time
remaining press the MODE button once to
view the MAX gas values screen.
The MAX figure represents the highest
gas value reading that has been recorded
by the instrument during the current
operating session or during the last 24
hours of operation. Press the MODE
button to display the Time in alarm screen
which shows the number of hours the
instrument has been in alarm.
If the Calibration reminder feature has
been enabled then Press the MODE
button to display the number of days until
calibration is due.
Otherwise the instrument will show the
first screen in the sequence.
Note: To reset the max readings level,
fresh air/zero calibrate the instrument
as discussed in section 3.3.