1. Overview
The Biosystems ToxiLtd hereafter referred
to as the ToxiLtd is a single sensor gas
detector that is designed to detect either
carbon monoxide (CO) or Hydrogen
Sulfide (H2S). The ToxiLtd’s sensor type
is shown on the front of the instrument
and is also shown on the display during
the startup sequence or during Gas
Indicator mode.
ToxiLtd gas detectors are disposable
instruments that are designed to operate
for 2 full years from the date of activation.
ToxiLtd models must be activated within 6
months of the manufacturing date. If the
instrument is not turned on by the “must
be activated by date”, the instrument will
activate itself and the 2-year countdown
will begin automatically.
Upon delivery, the screen of the ToxiLtd
will be blank except for the negated horn
icon.
If the “must be activated by date” has
passed and the instrument has activated
itself, the instrument will show three lines
on the screen with the negated horn icon.
Contact Sperian or your distributor if
three lines are displayed when you first
receive the detector.
The ToxiLtd is delivered with a pre-
installed lithium thionyl chloride (LiSOCl
2
)
battery capable of powering the
instrument for the full 2 years of the
instrument’s operating life unless the
instrument exceeds the allotted alarm
time. After the 2-year instrument life cycle
is up, the instrument will turn itself off.
1.1 Methods of sampling
The ToxiLtd may be used in diffusion
mode, or with the manual sample draw kit
that is available separately. In either
mode, the atmosphere must reach the
sensor for the instrument to register a
reading. In diffusion mode, the
atmosphere reaches the sensor by
diffusing through the sensor port on the
front of the instrument. Normal air
movements are enough to carry the
sample to the sensor. During remote
sampling, the gas sample is drawn into
the sensor compartment through the
probe assembly and a length of tubing.
See section 2.6 for more details on
sampling the atmosphere.
1.2 Sensors
The ToxiLtd uses an electrochemical toxic
gas sensor which has been designed to
minimize the effects of common
interfering gases. These sensors provide
accurate, dependable readings for gases
commonly encountered in industrial
applications.
1.2.1 Sensor ranges and resolutions.
Sensor Range Resolution
CO 0-1000 PPM 1 PPM
H
2
S 0 -200 PPM 1 PPM
1.2.2 F1: Missing sensor during
startup
If the ToxiLtd fails to detect the sensor
during startup, it will show “F 1” with the
caution symbol following the software
version screen at startup. This is covered
in greater detail in section 2.1.3.
1.3 Alarm logic
1.3.1 Gas alarms
When an alarm set point is exceeded a
loud audible alarm sounds, the bright red
LED alarm light flashes and the display
will indicate whether it is a warning alarm
or a more serious danger alarm. If the
ToxiLtd is equipped with the optional
vibrating alarm, it will also be activated
during an alarm condition. Instruments
have two ascending alarm set points.
The lower alarm set point is the warning
alarm, while the higher alarm set point is
the danger alarm.
Sensor Warning Alarm Danger Alarm
CO 35 PPM 100 PPM
H
2
S 10 PPM 20 PPM
Default Warning and Danger Alarm Settings