Reference
Manual
IQ6
Docking Station
for
PHD6
651 South Main Street
(800) 711-6776 (860) 344-1079
Fax (860) 344-1068
Part Number 13-324
Version 1.02
23AUG2010
Vista de pagina 0
1 2 3 4 5 6 ... 26 27

Indice de contenidos

Pagina 1

Reference Manual IQ6 Docking Station for PHD6 ‘ 651 South Main Street (800) 711-6776 (860) 344-1079 Fax (860) 344-

Pagina 2

92. Slide the open end of the tubing assembly over the regulator. 3. Connect the end of the tubing with the white quick disconnect fitting to the

Pagina 3 - Table of Contents

102.7.4 Multiple Cylinders to Multiple Docks The PHD6 includes 4 calibration gas inlet ports to provide calibration flexibility for gas detection pro

Pagina 4

11 The software will notify you when the process is complete. Click “Finish”. 2.8.2 Dock screens Once the connection is made, the dock will display

Pagina 5 - 1. Overview

12Indicators section of the IQ Ethernet Instructions Booklet. The current status of each dock is shown in the left column. When there are no detecto

Pagina 6 - 2. Installation

13of the screen. Once the dock configuration is uploaded, the “Set Config” button will be enabled. Once changes are made, click on the “Set Config”

Pagina 7

14 To darken or lighten the display on the dock, move the slider as appropriate. The dock will not show the new LCD setting until the “Set Config” b

Pagina 8

15 Once the gas type is selected, use the arrows to the right of the gas value input box to increase or decrease the value. If changes need to be ma

Pagina 9 - Gas Cylinder to the Dock(s)

16 To view the dock’s Display, click on “Get Current LCD Display”. Ethernet Controls For docks that are connected by Ethernet, there are three addit

Pagina 10

17appear on the right. Click on “View Session Data” to directly access session information for the instrument using the BioTrak Program. If the doc

Pagina 11

18must then be programmed with the same PIN and set to check for the code in the instruments. 3. Select the dock that will be used to write the PIN

Pagina 12 - 3. Software

1 THE IQ6 DOCKING STATION IS DESIGNED TO INTERFACE WITH PHD6 GAS DETECTORS. SPERIAN GAS DETECTORS HAVE BEEN DESIGNED FOR THE DETECTION OF DEFICIEN

Pagina 13

19 Enter your User ID and password again when the Security Login window is shown. A window will be shown to indicate that the dock configuration has

Pagina 14

204.1 Test Sequence Once the instrument has been recognized, the fresh air / zero calibration will begin automatically. Performing the fresh air

Pagina 15

21connected to the PC, the fault will also be indicated in the output box on the instrument tab in the IQ Express software. 4.3.1 No gas detected If

Pagina 16

225.2.1 Dock The Dock Tab under Tools / Options includes settings for the Dock Timeout and the Calibration Gas Lot Number. Each dock has its own set

Pagina 17

23large number of instruments are in use that are bump-tested daily though the IQ Express Dock. 5.2.4 Session Downloading The session downloading o

Pagina 18

24CAUTION Interfering with the flash upload may result in damage to the IQ6 Dock . Once the upload is completed, you’ll need to restart the PC for t

Pagina 19

25Appendix A: Calibration Frequency One of the most common questions that we are asked at Sperian Instrumentation is: “How often should I calibrate

Pagina 20 - 4. Dock Use

26Sperian Instrumentation Warranty Gas Detection Products General Sperian Protection Instrumentation, LLC (hereafter Sperian) warrants gas detectors,

Pagina 21 - 4.3 Test Failures

2Table of Contents Warnings and Cautions 3 A. Signal Words 3 1.Overview 4 1.1 Tests and record keeping 4 1.2 PC Connection 4 1.2.1 USB Connec

Pagina 22 - 5. Menu Items

3Optimize O2 Bump Test 15 Saving new settings 15 Dock LCD Display 15 Ethernet Controls 16 3.2.2 PHD6 Tab 16 3.2.2.1 Instrument Tab 17 General

Pagina 23

41. Overview The IQ6 Docking Station is an automatic calibration station for use with the Sperian PHD6 Gas Detectors. The IQ6 Docking Station automa

Pagina 24 - 6. Software Upgrades

5 30MB hard drive disk space.  USB Port *Sperian Instrumentation discourages multiple IQ Express Dock setups with PC’s running Windows 98. **Servi

Pagina 25

6the docks by connecting them to the PC’s USB port. If the instrument is to be connected via Ethernet, follow the Ethernet Instructions that came wi

Pagina 26 - Calibration Frequency

7 5. Click “Next” to continue. The Customer Information screen will be shown. 6. Enter the name of the user and the name of the organization. Th

Pagina 27 - Sensor Warranty Periods

8the instructions to create and specify the PostgreSQL database. 2.6 Power and Connectivity This section covers providing power to the IQ Express Do